Nov 21, 2026
6 min read
1 views
openFDA's device adverse event endpoint serves reports from FDA's MAUDE database (Manufacturer and User Facility Device Experience): medical device reports of suspected device-associated deaths, serious injuries and malfunctions, from mandatory reporters (manufacturers, importers and device user facilities) and voluntary reporters such as health care professionals, patients and consumers. It is free, updated weekly, and available as an API and as zipped JSON downloads. What it cannot show is stated by FDA itself: "a causal relationship cannot be established between product and reactions listed in a report," and "the incidence or prevalence of an event cannot be determined from this reporting system alone." FDA's MAUDE page adds that the data is not meant for evaluating adverse event rates, tracking rate changes over time or comparing rates across devices.
The openFDA device adverse event page says "The openFDA device adverse event API returns data from Manufacturer and User Facility Device Experience (MAUDE)," which it describes as "an FDA dataset that contains medical device adverse event reports submitted by mandatory reporters," alongside voluntary reports. FDA uses these medical device reports (MDRs) "of suspected device-associated deaths, serious injuries and malfunctions" to monitor device performance. The page lists the update frequency as weekly.
FDA's own MAUDE page, content current as of 12/08/2025, describes MAUDE as a database of MDRs of adverse events that also includes device and patient problems and patient demographic information. Its online search covers the last ten years; older reports are in the MDR Data Files, which the page says are updated monthly and cover voluntary reports since June 1993, user facility reports since 1991, distributor reports since 1993 and manufacturer reports since August 1996. The page says the MAUDE database itself is updated monthly.
API: the device event how-to page gives the endpoint https://api.fda.gov/device/event.json and search syntax of the form search=field:term. "The maximum limit allowed is 1000 for any single API call," and without a limit the API returns one matching record.
Rate limits: the openFDA authentication page gives, with no key, "240 requests per minute, per IP address. 1,000 requests per day, per IP address," and with a key, "240 requests per minute, per key. 120,000 requests per day, per key." "The key is free of charge."
Downloads: the openFDA downloads page says files are zipped JSON "in exactly the same JSON format that query results follow," listed in a manifest, with large categories "broken up into many small parts." It warns that "you cannot simply download" new files to keep a downloaded copy up to date.
The openFDA terms say that, unless otherwise noted, openFDA content and data "is public domain and made available with a Creative Commons CC0 1.0 Universal dedication." The same page says "Some data on openFDA may not be public domain, such as copies of copyrightable works," and that works where CC0 does not apply "will be clearly marked by a warning in the relevant documentation." It says openFDA includes content supplied by The GMDN Agency and used under licence, and that extracting GMDN content to build commercial services, such as alternative categorisation, alternative term assignment or mapping to other ontologies, or for AI training, requires a licence first. The platform is provided 'as is' and on an 'as-available' basis, in the page's words, and FDA "reserves the right to stop providing openFDA at any time for any reason in its sole discretion." The terms page as read shows no last-updated date. The device event pages as read do not say which fields, if any, carry GMDN content.
From the openFDA device adverse event page:
Data quality: passive surveillance brings "potential submission of incomplete, inaccurate, untimely, unverified, or biased data," and reports "do not undergo extensive validation or verification."
Causation: "a causal relationship cannot be established between product and reactions listed in a report."
Incidence: "the incidence or prevalence of an event cannot be determined from this reporting system alone," with "potential under-reporting of events and lack of information about frequency of device use."
Denominators: reports "represent a small percentage of total usage numbers of a product."
Trends: "Increases in the total number of adverse events is likely caused by improved reporting."
UDI: "Not all MDRs provide UDI data, and the current data is not validated."
Clinical use: reports "should not be the sole source of information for clinical decision making."
From FDA's MAUDE page:
Rates: MDR data is not meant to evaluate adverse event rates, track changes in rates over time or compare rates across devices.
Death in the text is not a death report: a report mentioning death is not classified as a death unless the reporter believes the device was or may have been a factor.
Not an admission: under 21 CFR 803.16, submitting or releasing a report is not an admission that the device or the reporter caused the event.
Two coverage statements on one page. The openFDA device event page lists "2009 to null" in its key facts and "from about 1992 to the present" in its about text. Check the dates actually returned before stating coverage.
Weekly API, monthly MAUDE. openFDA lists weekly updates; FDA's MAUDE page says MAUDE and the MDR Data Files are updated monthly.
Reports change. FDA's MAUDE page says MAUDE shows the most recent information, so earlier details can be overwritten or appended by follow-up reports.
One report, many events. Some reports summarise multiple events, including Voluntary Malfunction Summary Reporting and literature-based summaries, flagged by the "Summary Report (Y/N)" field.
Missing reports. Reports submitted under 21 CFR 803.19 exemptions, variances or alternatives may be missing if they predate June 2019; real-world data exemptions carry "RWD" identifiers.
Redactions. Trade secret or confidential business information appears as "(b)(4)" and personal or medical information as "(b)(6)."
Partial downloads mislead. The downloads page says new files alone do not keep a local copy current.
For recall data from the same service, see FDA medical device recall data on openFDA.
Glasp has no published note on openFDA device adverse event data.
No. FDA says incidence and prevalence cannot be determined from the reporting system alone and that a causal relationship cannot be established from a report.
openFDA lists weekly updates for the API. FDA's MAUDE page says MAUDE and the MDR Data Files are updated monthly.
The API and downloads are free. openFDA's terms dedicate its content to CC0 unless otherwise noted, with marked exceptions and a licence requirement for certain uses of GMDN content.
The openFDA page says increases in total adverse events are "likely caused by improved reporting," and FDA's MAUDE page says MDR data is not meant to track rate changes over time.
Compiled by Glasp from public pages read on November 22, 2026. Terms, limits and coverage are quoted as they appeared that day and may change; check FDA's and openFDA's pages before relying on them. This page is not medical advice.